ICP-MS法测试地质样品的前处理研究
赵文志;张元;杜少文;张兵兵;李树坤
【摘 要】In this paper,the standard substance ESS2,ESS3 and ESS4 were
simultaneously digested by electric heating plate driving acid
method,airtight digestion method and microwave digestion method,The
contents of 7 elements,Bi,Cd,Co,Cu,Ni,Pb and Zn,were determined by ICP-
MS,and the digestion effects of the three methods are
result shows:The detection limits of three digestion methods were 0.008~
0.391,0.014~0.353,0.011~0.347μg·g-1,all of which had higher detection
capability;The accuracy of the determination of 7 elements was less than
0.1,and the relative standard deviation was lower than 10%,so it has a
good detection limit,accuracy and precision.%本文采用电热板消解法、密闭
消解法及微波消解法消解标准物质ESS2、ESS3和ESS4,用ICP-MS分别测定Bi、
Cd、Co、Cu、Ni、Pb和Zn 7种元素的含量,对3种方法的消解效果进行比较.结
果表明:3种方法的检出限分别为0.008 ~ 0.391,0.014 ~ 0.353,0.011 ~
0.347μg·g-1,均具有较高检出能力;测定7种元素的准确度结果均小于0.1,相对标
准偏差均低于10%,具有很好的检出限、准确度及精密度.
【期刊名称】《化学工程师》
【年(卷),期】2018(032)002
【总页数】4页(P25-28)
【关键词】ICP-MS;电热板赶酸;密闭消解;微波消解
【作 者】赵文志;张元;杜少文;张兵兵;李树坤
【作者单位】中国人民武装警察部队黄金第三支队化验室,黑龙江哈尔滨150039;
中国人民武装警察部队黄金第三支队化验室,黑龙江哈尔滨150039;中国人民武装
警察部队黄金第三支队化验室,黑龙江哈尔滨150039;中国人民武装警察部队黄金
第三支队化验室,黑龙江哈尔滨150039;中国人民武装警察部队黄金第三支队化验
室,黑龙江哈尔滨150039
【正文语种】中 文
【中图分类】O657.63
地质样品具有多样性的特点,其中耐酸性能较好,一般很难对其消解,地质样品的
测试需要和实际样品的特质相结合,从而选择合适的测试方法,并且能够在一定程
度上减少试验过程中所产生的污染物,弱化其对生态环境的破坏,还可以减少成本
的投入,从而更加高效、便捷的达到测试目的。本文采用电热板消解、密闭消解及
微波消解3种前处理方法消解国家标准物质E SS系列 [1-5],用ICP-MS检测Bi、
Cd、Co、Cu、Ni、Pb、Zn 的含量,对比其检出限、准确度及精密度,比较3种
消解方法对测定结果的影响,微波消解的准确度和精密度最好,但一次消解的样品
数较少,不能满足大批次测试的要求,电热板消解重现性相对较差,但一次消解样
品数量多,适用于大批量样品分析;本文验证了电热板消解在实际工作中的可行性,
电热板消解测定的结果可靠,数据可以采用,能满足分析质量的要求。
1 实验部分
1.1 主要仪器和试剂
X Serie sⅡICP-MS(T hermofi s her,美国);超纯水(18.2MΩ·cm-1);HC
l、HN O3均为优级纯,高纯 A r(质量分数>99.99%);
仪器调试组合溶液ρ(B)=1n g·m L-1,介质2%HN O3,B=Ba,Be,Bi,Ce,
Co,In,Li,Ni,Pb,U;
标准溶液配制:本试验均采用与试样同时制备的国家一级标准物质 G SS27,G
SS4,G SS22,G S D-5a,G S D-3a,G S D-9,G S D-14,G S D-10作为标
准物质溶液。
1.2 仪器工作参数的选择
仪器工作参数主要考察了射频功率、等离子气流速、雾化气流速、辅助气流速、获
取时间、采样深度等因素对强度的影响。按照能够使检测信号强度最大,相对标准
偏差(RS D)最小的原则进行优化。在ICP-MS测定中,射频功率选为
1300W[6]。最佳仪器参数见表1。
表1 ICP-MS仪器工作条件(ICP-MSXSeriesⅡ)Tab.1 Working conditions of
ICP-MS(ICP-MSXSeriesⅡ)仪器工作参数 数据获取参数I C P功率/W 1300
扫描方式 跳峰冷却气(A r)流量/L·min-1 13.5 点数/质量峰 3辅助器(A r)
流量/L·min-1 0.75 停留时间m s/点 10雾化器(A r)流量/L·min-1 0.85 扫
描次数/样品 50取样锥(Ni)孔径/mm 1.0 总测量时间/s 60截取锥(Ni)孔
径 /mm 0.7
1.3 样品前处理
1.3.1 电热板消解基本步骤 称取样品0.1000g,置于聚四氟乙烯坩埚中,分别加
入5m LHN O3、5m LH F和1m L HC l O4,将坩埚置于电热板升温至160℃加
热1h,调节温度至210℃加热6h,然后将温度降至150℃,趁热加入 5m L 王水
(1+1),加热 15min,等至溶液清亮后冷却,将溶液倒入25m L比色管中,用
超纯水定容。移取1m L于比色管中,用3%的HN O3稀释至10m L,摇匀后上
机测试。随同试验做2份空白[7-9]。
1.3.2 密闭消解操作步骤 称取样品0.1000g置于密闭消解罐中的聚四氟乙烯内衬
中,分别加入4m L HN O3和4m LHF,将其置于控温箱中调节温度至180℃加
热6h。冷却后,将内衬置于电热板上蒸干,用1m LHN O3浸提,将溶液倒入
25m L比色管中,用超纯水定容,移取1m L于比色管中,用3%的硝酸稀释至
10m L,摇匀后上机测试。随同试验做2份空白[10,11]。
1.3.3 微波消解法操作步骤 称取样品0.1000g于消解罐中,分别加入3m LHN
O3、2m LH2O2及1m LHF,预消解1h后将其放入至微波消解仪内,调节温度
至120℃消解5min,升温至180℃消解30min。消解完毕后赶酸、用超纯水定容
至25m L。移取1m L于比色管中,用3%的硝酸稀释至10m L,摇匀后上机测试。
随同试验做2份空白[12,13]。
2 结果与讨论
2.1 方法检出限比较
在选定的实验条件下,对试剂空白溶液连续测定12次,计算其标准偏差,以3倍
标准偏差计算待测元素的检出限,其结果见表2。
表2 3种消解方式下ICP-MS测定的方法检出限Tab.2 Detection limit of ICP-
MS by three digestion methods?元CCCPZN B素 bn du o i i
0000000电标偏.......0001100热86 00232准差5230563板0703037
赶酸检(消出L0003321 D1069759)解限5891591
0000000标偏.......0001110密021014 0准差6417317闭3707737 消
检(0000000解.......出L0003331 D1165134 )限 9433143 0000000
标偏.......0001010微14 00209准差6348855波7770777 消检
(0000000解.......出L0003231 D2172943)限0144677
由表2可见,3种前处理方法的测定结果分别为 0.008~0.391,0.014~0.353,
0.011~0.347μg·g-1;其中电热板消解的Bi、Cd和Pb检出限最低,密闭消解的
Co检出限最低,微波消解的Cu、Ni和Zn检出限最低;同时比较各元素的检出
限,3种前处理均是Cd的检出限最低,电热板消解和密闭消解时Cu的检出限最
高,微波消解时Pb的检出限最高。3种前处理方法均具有较好的检出能力。
2.2 准确度和精密度比较
对3种前处理方法的标准样品溶液进行12重复测定,测定结果见表3。
表3 3种消解方式下ICP-MS测定的准确度和精密度Tab.3 Accuracy and
precision of ICP-MS determination under three digestion methodsICP-MS
测定值元素 标样 标准值 电热板赶酸消解 密闭消解 微波消解平均值 △logC
RS D% 平均值 △logC RS D% 平均值 △logC RS D%Bi G SS-2 0.380 0.383
0.003 4.777 0.384 0.005 3.064 0.383 0.004 3.039 G SS-3 0.170 0.177
0.017 7.663 0.174 0.011 5.296 0.173 0.006 5.094 G SS-4 1.040 1.013 -
0.012 4.082 1.013 -0.012 2.360 1.033 -0.003 2.909 Cd G SS-2 0.071
0.067 -0.024 2.806 0.068 -0.020 2.485 0.070 -0.007 2.536 G SS-3 0.060
0.055 -0.038 4.049 0.055 -0.036 3.749 0.057 -0.026 1.946 G SS-4 0.350
0.308 -0.055 3.621 0.308 -0.056 6.461 0.328 -0.029 4.892 Co G SS-2
8.700 8.558 -0.007 5.931 8.908 0.010 5.401 8.850 0.007 4.790 G SS-3
5.500 5.200 -0.024 9.238 5.417 -0.007 6.571 5.308 -0.015 6.461 G SS-4
22.000 22.692 0.013 9.659 21.767 -0.005 7.802 21.867 -0.003 5.238
Cu G SS-2 16.300 16.208 -0.002 5.421 16.367 0.002 4.675 16.325
0.001 3.530 G SS-3 11.400 11.373 -0.001 9.799 11.599 0.008 7.579
11.550 0.006 6.55 G SS-4 40.000 40.325 0.004 5.931 40.883 0.009
3.782 40.308 0.003 2.768 Ni G SS-2 19.400 20.267 0.019 7.164 19.983
0.013 5.306 20.250 0.019 4.603 G SS-3 12.000 12.300 0.011 9.341
12.342 0.012 5.702 11.900 -0.004 6.02 G SS-4 64.000 62.558 -0.010
6.623 62.292 -0.012 3.275 64.508 0.003 2.276 Pb G SS-2 20.000
19.242 -0.017 4.320 21.325 0.028 6.658 21.367 0.029 6.827 G SS-3
26.000 24.733 -0.022 3.521 26.125 0.002 6.093 26.508 0.008 4.65 G
SS-4 58.000 54.983 -0.023 3.359 59.450 0.011 4.51 57.325 -0.005
4.338 Zn G SS-2 42.000 40.892 -0.012 4.290 41.900 -0.001 3.513
41.750 -0.003 3.938 G SS-3 31.000 31.867 0.012 6.366 31.900 0.012
4.836 31.050 0.001 4.035 G SS-4 210.000 206.283 -0.008 5.255
211.208 0.002 4.49 210.57 0.001 4.007
测定结果表明,3种方法的测定值均满足要求,元素的化学性质不同,采用不同消
解方法时,测试结果有所差异。密闭消解和微波消解测定的Cd、Cu值整体偏低,
可能是由于消解时元素的溶出率较高,导致其他元素对Cd、Cu的测定产生较大
干扰。电热板消解的Pb值偏低,可能是由于消解时形成的PbCl2挥发,导致测
定结果偏低[14]。
本文采用△logC来衡量元素测定的准确度,3种前处理方法的测定值和标准值基本
保持一致,其中微波消解的准确度最好,其次是密闭消解,其结果均小于0.1,这
表明3种方法均适用于上述7种元素的测定。3种消解方法测定上述7种元素的
相对标准偏差均低于10%,均具有较好的精密度,总体上微波消解要优于密闭消
解和电热板消解,其次为密闭消解。但电热板消解测定上述7种元素其精密度仍
满足规范要求。
2.3 方法可操作性比较
微波消解具有消解完全,消解时间短、加热均匀的特点,但是一次消解的样品数量
较少,不能满足大批次测试的要求[15];密闭消解具有一定的危险性,转移时酸的
残留量较大,对人的危害较大。电热板消解虽然重现性相对较差,但是一次消解样
品数量多,消解成本和仪器设备要求要低于微波消解和密闭消解,能够满足大批量
样品分析。
为了验证电热板消解在实际工作中的可行性,对40个批次的样品插入标准物质G
SS22作为监控样作出质量监控图。以50个样品作为一批,插入至少1个标准物
质作为监控样作出质量监控图,每批次带有监控样的测定值如果均落在上下控制线
之内,则表明测定结果可信,数据可用;如果结果落在上下控制线之外,则表明测
定值不可信,数据不可用。由图1可以看出,监控样的测定值大多数落在中心线
附近并且所有数据都落在控制线之内,说明电热板消解能够满足测试质量的要求。
图1 七种元素的质量控制图Fig.1 Quality control chart of seven elements
3 结论
在本文中,采用了电热板消解、密封消解及微波消解 3 种方法测定了 Bi、Cd、Co、
Cu、Ni、Pb 和 Zn 7种元素的含量,3种方法消解标准样品的测定值与标准值基
本一致,准确度结果均小于0.1,相对标准偏差均低于10%,具有很好的检出限、
准确度和精密度;验证了电热板消解在实际工作中的可行性,通过质量监控图得出
结论,电热板消解测定的结果可信,数据可用,能够满足分析质量的要求。
参考文献
[1] 何红蓼,李冰,韩丽荣,等.封闭压力酸溶ICP-MS法测定地质样品中 47 个
元素的评价[J].分析实验室,2002,21(5):8212.
[2] 汤志勇.岩石矿物分析[J].分析试验室,2012(12):108-124.
[3] Freitas,R.,Ramos, N. L.,Batista, B. L., & Barbosa Jr.,F. Toxic and
essential elements in conventional and home-produced eggs by ICP-MS
Additives& B,Surveillance,2013,6(1):
30-35.
[4] 周国华.多目标区域地球化学调查:分析测试面临的机遇和挑战[J].岩矿测
试,2010,(6):296-300.
[5] 叶轶凡.岩石矿物分析鉴定浅析[J].科技资讯,2011,(14).
[6] Guo,X.M.;Sturgeon,R.E.;Mester,Z.;GardneG.J. Photoehemical
alkylation of inorganic Part fication of volatile [J].J.
. SPectrom.2005,20:702-708.
[7] Albanese,S.,De Vivo,B.,Lima,.,&Cicchella,D. Geochemical
background and baseline values of toxic elements in stream sediments of
Campania region(Italy)[J].Journal of Geochemical Exploration,2007,
93:21-34.
[8] 尹明,李家熙,等.岩石矿物分析(第四版)[M].地质出版社,2011.2.
[9] Martuzzi,M.,Mitis,F.,Bianchi,F.,Minichilli,F.,Comba,P.,
&Fazzo,L. Cancer mortality and congenital anomalies in a region of
Italy with intense environmental pressure due to waste[J].Occupational
and Environmental Medicine,2009,66:725-732.
[10] Khan,K.,&Naeem,aneous determination of
accumulated hazardous metals in hen's eggs by atomic absorption
spectroscopy
[J]. Journal of Applied Sciences,2006,6(1):198-201.
[11] Rappel C,Schauml ffel role of sulfur and sulfur isotope
dilution analysis in quantitative protein analysis[J].Anal Bioanal
Chem,2008,390(2):605-615.
[12] 李月芳,姚檀栋,王宁练,等.电感耦合等离子体质谱直接测定冰芯样品中
痕量铅[J].岩矿测试 ,2001 ,20 7(1):11214.
[13] 李冰,杨红霞.电感耦合等离子体质谱原理和应用[M].北京:地质出版社,
2005.
[14] Landaud S,Helinck S,Bonnarme P. Formation of volatile sulfur
compounds and metabolismof methionine and other sulfur
compounds in fermented food[J].Applied microbiology and
biotechnology,2008,77(6):1191-1205.
[15] 胡圣虹,林守麟,刘勇胜,等.等离子体质谱法测定地质样品中痕量稀土元
素的基体效应及多原子离子干扰的校正研究[J].高等学校化学学报,2000,21
7(3):3682372.
发布者:admin,转转请注明出处:http://www.yc00.com/num/1711498550a1915921.html
评论列表(0条)